Table 2 Procedural data of the study cohort

From: Multipolar pacing by cardiac resynchronization therapy with a defibrillators treatment in type 2 diabetes mellitus failing heart patients: impact on responders rate, and clinical outcomes

Parameters

Quadripolar group (n 99)

Bipolar group (n 96)

p value

Skin to skin time

183 ± 113

179 ± 116

n.s

Fluoro time

17 ± 4.7

18 ± 5.6

n.s

CS cannulation time

14 ± 5.4

15 ± 4.7

n.s

CS target vessel

 Anterior, n (%)

3 (3)

3 (3)

n.s

 Anterolateral, n (%)

13 (13)

11 (12)

n.s

 Lateral, n (%)

51 (51)

52 (54)

n.s

 Postero-lateral, n (%)

28 (28)

25 (26)

n.s

 Posterior, n (%)

4 (4)

5 (5)

n.s

LV lead parameters

 Impedance thresholds

761 ± 182

685 ± 196

n.s

 Pacing thresholds

0.5 ± 0.4

0.5 ± 0.3

n.s

 Sensing thresholds

14 ± 7.2

13.4 ± 7.3

n.s

RA lead parameters

 Impedance thresholds

438 ± 164

452 ± 133

n.s

 Pacing thresholds

0.4 ± 0.2

0.5 ± 0.3

n.s

 Sensing thresholds

2.1 ± 1.7

2.3 ± 1.8

n.s

RV lead parameters

 Impedance thresholds

581 ± 151

577 ± 163

n.s

 Shock impedance thresholds

73 ± 15

76 ± 14

n.s

 Pacing thresholds

0.5 ± 0.45

0.43 ± 0.35

n.s

 Sensing thresholds

19 ± 6

18 ± 5

n.s

  1. In this table we reported procedural data as functionality parameters of devices leads (left ventricle, right atrium, and right ventricle leads), and procedural times at implant (shin to skin time, fluoro time, CS target vessel, and CS cannulation time) comparing multipolar vs bipolar group of patients. Impedance thresholds are expressed in Ohm. Pacing thresholds are expressed in millivolts for milliseconds. Sensing thresholds are expressed in millivolts. Procedural times (skin to skin time, fluoro time, and CS cannulation time) are expressed in minutes. The “n.s” was for statistical not significant (p value >0.05)
  2. CS coronary sinus, LV left ventricle, RA right atrium, RV right ventricle